How to maintain and repair ultrasonic cleaning machines

Time:2014-10-14

The cleaning machine consists of an ultrasonic generator and an ultrasonic transducer. The ultrasonic transducer is a sandwich type transducer made of piezoelectric ceramic material, which generates mechanical vibration under the action of an alternating electric field.

 

Common problems with ultrasonic cleaning machine generators include:

⑴ Turn on the power switch of the ultrasonic cleaning machine, but the indicator light does not light up. When repairing this situation, it is necessary to check whether the power switch is intact and whether the leakage switch is closed. If the switch is intact, then check if the fuse is overloaded and blown, which can basically solve the problem.

After turning on the power switch, the indicator light is on, but there is no ultrasonic output. This situation is quite complex. When repairing, first check if the connection plug between the transducer and the ultrasonic power board is loose, and then check if the fuse is blown. If everything is normal, it is possible that there is an internal fault in the ultrasonic power generator. Use a multimeter to open the power cord and check if all the live wires are connected. After troubleshooting the generator, check if the ultrasonic transducer is burnt out and if it needs to be replaced.

The DC fuse of the ultrasonic cleaning machine is blown. It may be due to the burning of the rectifier bridge stack or power tube, or it may be due to aging of the transducer and unstable current, all of which can cause power generator failure. Pay more attention during maintenance.

After turning on the power switch, the ultrasonic cleaning machine produces ultrasonic output, but the cleaning effect is not as ideal as before. This situation is mostly caused by improper use, such as improper cleaning liquid level in the cleaning tank, which is too high or too low; The ultrasonic frequency coordination was not adjusted properly; The liquid temperature in the cleaning tank is too high or too low; Improper selection or insufficient amount of cleaning solution. Of course, it cannot be ruled out that after the transducer is debonded, most of the ultrasonic energy cannot be transmitted to the cleaning solution. In this case, it should be promptly sent back to the manufacturer for repair to avoid burning the oscillator.

Common issues with ultrasonic transducers:

(1) "Ultrasonic transducer is damp". Generally, a megohmmeter is used to check the plug connected to the transducer. By checking the insulation resistance value between the positive and negative poles of the transducer, it can be determined. It is generally required that the insulation resistance be greater than 30 megaohms or more. If the insulation resistance value cannot be reached, it is likely that the transducer is damp. The maintenance method is to place the entire transducer (excluding the spray coated shell) in an oven set at around 100 ℃, dry it for three hours, or use a hair dryer to remove moisture until the resistance value is normal. The transducer oscillator ignites, causing the ceramic material to shatter. During maintenance, a combination of visual inspection and a megohmmeter can be used for inspection. Generally, as an emergency response measure, individual damaged transducers can be disconnected without affecting the normal use of other transducers We know that most manufacturers use adhesive to fix transducers, but ultrasonic cleaning machines may experience debonding due to vibration after long-term use. Some manufacturers also use the method of bonding and tightening with screws, which generally does not cause delamination. Due to the action of the screws, the oscillator will not fall off the vibration surface after delamination. The general judgment method is to gently shake the tail of the oscillator by hand and carefully observe the adhesive situation on the vibration surface to make a judgment.

After the detachment of the general oscillator, the output power of the ultrasonic power supply is normal. However, due to poor connection between the oscillator and the vibration surface, the vibration effect of the vibration surface is not good. After a long time, due to the inability to release energy, it is likely to burn out the oscillator. The detachment of the oscillator is quite troublesome for users to repair, and in general, it can only be sent back to the manufacturer for repair. An effective way to avoid detachment of the oscillator is to avoid hitting the vibrating surface during normal use.

(4) "Vibration surface perforation": Generally, ultrasonic transducers may experience vibration surface perforation after being used at full load for several years. This is due to long-term high-frequency vibration fatigue of the stainless steel plate on the vibration surface. Vibration surface perforation indicates that the service life of the transducer has expired and can only be replaced for maintenance.

   

 

 

 

Disclaimer:The authenticity of this information has not been verified by our website and is for your reference only. The purpose is to convey more information; If there is any error, error, or infringement, please contact us via email( chinavta@szxwds.com )We would greatly appreciate it if you could promptly notify us of the error, error, or infringement, and we will promptly make corrections or deletions.

 

主站蜘蛛池模板: 国产aⅴ无 | 97免费无 | 伊人焦久综合影院每日更新 | 男人的天堂!天天色亚洲 | 极品吹潮视频 | 粗大的内捧| 日韩免费网页版视频 | 另类专区 | 日本邪恶网站 | 成年片色大黄全免费网站观看 | 欧美精品成人三级在线观看 | 亚洲欧美日本国产专区一区 | 日本一区二区三区四区不卡 | 国产丰满老熟女重口对白 | 人片在线观看 | 牛牛在线视频 | 吉吉影音先| 人摸人人 | 国产激情视频一区二区三区 | 日韩精品免费高清视频在线 | 欧美性爱| 国产精品天天在线午夜更新 | 门国产乱 | 不打码在线tv播放观看 | 国产尤物a∨视频在线观看 韩国三级私人教练 | 日韩视频网 | 99久免费精品视频在线观看 | 国产专区一区 | 国产一级二级三级精品视频 | 99热国产在线播放只有精品 | 国产精品尤物 | 痴汉系列在线观看 | 亚洲福利一区福利三区 | 亚洲激情网站 | 国产精品中文字幕制服 | hd护士18 | 日韩免费观看一区国产 | 日韩在线免费视频 | 香蕉一区二区在线观看 | 欧美日韩1区2区国产线 | 国产国产在线观看 |